TMQ.CVK pulse vacuum
autoclave is a small-sized automatic high pressure and temperature
sterilization equipment which takes pressure steam as medium. It is
suitable for sterilizing articles likes dressings,utensils, Class B
cavum equipment, medical catheter that used in stomatology,dentistry,
surgical, laboratory. |
 |
| ¡¡ |
Main feature and characters |
| ¡¡ |
¡öIt is controlled by
macro-processor, touch screen, and SCM is adopted as main control.Those
makes the system stable and perfect.
¡öTemperature, pressure, time, type are shown
digitally , and there are states indicator.
¡öThere are three set sterilize program, 2 test
program for the Users.
¡öIt has BD test program for dressing package steam
penetrability inspection.
¡öIt has vacuum test program, vacuum leakage is
less than 0.13Kpa/min.
¡öVacuum pump which imported from Germany is
equipped, vacuum degree could reach -0.08MPa(-0.8Bar)after three times
pulse. The air in the clearance of dressing, sterilizing bag and dentist
hand device can be pumped out to reach perfect steam saturation.
¡öPulse behind vacuum drying system, the vacuum can
reach -0.08Mpa (-0.8Bar), and the residue humidity of instrument after
drying will not exceed 0.2%.
¡öAutomatic electron safety lock ensures the door
be locked when there¡¯s automatic warning,faults and no power.
¡ö0.22 ¦Ì m bacterium air filter of air after
vacuum, re-pollution can be avoided.
¡öFaults could be inspected automatically, and
faults code could be displayed digitally.
¡öBuilt-in water tank, pure water and dirty water
could be separated.
¡öPrinter could be equipped according to users
requirements.
¡öBuzzer gives alarm when all the cycles are ended. |
| ¡¡ |
Safety device |
| ¡¡ |
¡öThe system will auto
cut off power when temperature over the rated temperature.
¡öDoor safety device: electric lock makes the
sterilization could be run under conditions of the door is closed
correctly, if the chamber ahs pressure the door can not be opened.
¡ö Lower water level alarm device: when water level
in the tank is
lower than required, the power will be cut off automatically,
meanwhile alarm signals will be sent.
¡ö When chamber pressure is over rated value the
safety valve could be opened automatically.
¡ö Electron safety device: the DC control loop with
over pressure
protection, the AC loop has breaker. |
| ¡¡ |
Technique Parameters |
| ¡¡ |
Design pressure: -0.1 ¡«
0.26MPa
Rated work pressure: 0.23Mpa
Limit vacuum degree: -0.08MPa
Pressure control precision: 1KPa
Sterilization temperature: 105 ¡« 136 ¡æ
temperature control precision0.1 ¡æ
Maximum temperature: 136 ¡æ
Temperature uniform: ¡Ü¡À0.5 ¡æ
Chamber material : Stainless steel SUS304 , 3 pieces trays made of
aluminum |
| ¡¡ |
|
Model |
Capacity |
Chamber Dimension
(¦µ¡ÁL) |
Package
(L¡ÁW¡ÁH) |
Net
Weight |
Voltage |
Frequency |
Power |
| TMQ.CVK-2540S |
23L |
250*440mm |
720*520*440mm |
58kg |
220V |
50Hz |
2.5Kw |
| TMQ.CVK-2540B |
23L |
250*440mm |
720*520*440mm |
58kg |
220V |
50Hz |
2.5Kw |
| ¡¡ |
¡¡ |
¡¡ |
¡¡ |
¡¡ |
¡¡ |
¡¡ |
¡¡ |
|
¡¡ |