TMQ.CV series table top Autoclave
¡¡

   Contact Us      

¡¡ Hot air used as medium, GMQ full-auto-matic hot air sterilizer is a small-sized quick sterilizing equipment. It is suitable to be used sterilizing the loads that can stand hot air such as surgical instruments, dentistry instruments, glass utensil, and paraffin etc.in operation room, stomatology depart ment opahthalmic department and biology medical research department.
¡¡ Main feature and characters
¡¡

¡öIt is controlled by macro-processor, touch screen, and SCM is adopted as main control.Those makes the system stable and perfect.
¡öTemperature, pressure, time, type are shown digitally , and there are states indicator.
¡öThere are three set sterilize program, 2 test program for the Users.
¡öIt has BD test program for dressing package steam penetrability inspection.
¡öIt has vacuum test program, vacuum leakage is less than 0.13Kpa/min.
¡öVacuum pump which imported from Germany is equipped, vacuum degree could reach -0.08MPa(-0.8Bar)after three times pulse. The air in the clearance of dressing, sterilizing bag and dentist hand device can be pumped out to reach perfect steam saturation.
¡öPulse behind vacuum drying system, the vacuum can reach -0.08Mpa (-0.8Bar), and the residue humidity of instrument after drying will not exceed 0.2%.
¡öAutomatic electron safety lock ensures the door be locked when there¡¯s automatic warning,faults and no power.
¡ö0.22 ¦Ì m bacterium air filter of air after vacuum, re-pollution can be avoided.
¡öFaults could be inspected automatically, and faults code could be displayed digitally.
¡öBuilt-in water tank, pure water and dirty water could be separated.
¡öPrinter could be equipped according to users requirements.

¡¡ Safety device
¡¡

¡öThe system will auto cut off power when temperature over the rated temperature.
¡öDoor safety device: electric lock makes the sterilization could be run under conditions of the door is closed correctly, if the chamber ahs pressure the door can not be opened.
¡ö Lower water level alarm device: when water level in the tank is
lower than required, the power will be cut off automatically,
meanwhile alarm signals will be sent.
¡ö When chamber pressure is over rated value the safety valve could be opened automatically.
¡ö Electron safety device: the DC control loop with over pressure
protection, the AC loop has breaker.
¡ö Overheat protection: the system will shut off the heating power if temperature exceeds setting value

¡¡ Technique Parameters
¡¡ Design pressure: -0.1 ¡« 0.26MPa
Rated work pressure: 0.23Mpa
Limit vacuum degree: -0.08MPa
Pressure control precision: 1KPa
Sterilization temperature: 105 ¡« 136 ¡æ
temperature control precision0.1 ¡æ
Maximum temperature: 136 ¡æ
Temperature uniform: ¡Ü¡À0.5 ¡æ
Chamber material : Stainless steel SUS304 , 3 pieces trays made of aluminum
¡¡
Model Capacity Chamber Dimension
(¦µ¡ÁL)
Package
(L¡ÁW¡ÁH)
Net
Weight
Voltage Frequency Power
TMQ.CV-3850 60L 380*500mm 985*736*676mm 105kg 380V 50Hz 5.7Kw
TMQ.CV-3870 80L 380*700mm 985*736*676mm 120kg 380V 50Hz 5.7Kw
¡¡ ¡¡ ¡¡ ¡¡ ¡¡ ¡¡ ¡¡ ¡¡
¡¡
¡¡
¡¡