¡¡ TMQ.R series automatic table top autoclave of class N
¡¡
¡¡ Micro-computer control, touch button
¡¡

¡ö Heating, sterilization, steam exhaust and drying process are automatically
¡ö The process, temperature and time can be displayed with LCD,the faults can be alarmed and displayed
¡ö The function exhaust with positive pressure pulse can eliminate cool air in chamber to ensure the saturation steam
¡ö Buzzer can speak when the process finished
¡ö Steam and water built-in circulate, don¡¯t need water source and don¡¯t exhaust steam
¡¡

¡¡ Technique Parameters
¡¡ Design pressure:  0.26MPa
Design temperature: 150¡æ
Rated work pressure: 0.23Mpa
Sterilization temperature: 105 ¡« 136 ¡æ
temperature control precision0.1 ¡æ
Maximum temperature: 136 ¡æ
Temperature uniform: ¡Ü¡À0.5 ¡æ
Chamber material : Stainless steel SUS304
¡¡ Safety device
¡¡

¡ö Safety valve can be open if pressure exceeds setting value
¡ö Door interlock: heating for steam only starts when the door closes. The door can be locked if there is pressure in chamber
¡ö Overheat protection: the system will shut off the heating power if temperature exceeds setting value, alarm with sound and light
¡ö The program will be stopped if water level lower than the low limit, alarm with sound and light
¡ö The system will shut down power at short circuit
¡ö The system will shut down power at electric leakage
¡ö The system will shut down power at over-high voltage

¡¡
Model Capacity Chamber Dimension
(¦µ¡ÁL)
Package
(L¡ÁW¡ÁH)
Net
Weight
Voltage Frequency Power
TMQ.R-2540 23L 250*440mm 600*490*370mm 38kg 220V 50Hz 2.2Kw
TMQ.R-3250 45L 315*575mm 690*555*450mm 70kg 220V 50Hz 2.85Kw
¡¡ ¡¡ ¡¡ ¡¡ ¡¡ ¡¡ ¡¡ ¡¡
¡¡
¡¡
¡¡